¡¤ 8947Àº ´ÙÀ½ÀÇ ¾îÇø®ÄÉÀ̼ÇÀ» À§Çؼ »ç¿ëÇÒ ¼ö ÀÖ´Â:ÃøÁ¤ÇÏ´Â ¿£Áø Áøµ¿, ¼¿½Ä ÀüÈ, CDÇ÷¹À̾î Áøµ¿ ½ÃÇèÁß¿¡ ¹ÙÀ̺극ÀÌÅÍÀÇ Áøµ¿À» ÃøÁ¤ÇÏ´Â °ü°è Áøµ¿À» ÃøÁ¤ÇÏ´Â ÆßÇÁ Áøµ¿À» ÃøÁ¤ÇÏ´Â ºê·¹ÀÌÅ© Áøµ¿ÀÇ ÃøÁ¤...¶Ç ´Ù¸¥ ŸÀÔÀÇ Áøµ¿ ³»±¸¼º Å×½ºÆ® |
|
|
|
Input |
Number of channels: 2, Measurement objects can be selected individually for each channel. Full isolation between inputs, and between inputs and recorder. Common GND for voltage input and charge input channels. Voltage and pre-amplifier internal inputs: BNC terminals
Charge input: miniature connector (#10-32 UNF) |
Suitable converters |
Charge input: piezoelectric charge output acceleration pickup sensors Internal pre-amplifier input: acceleration pickup sensors with built-in pre-amplifier |
Measurement ranges
Charge input (miniature connector) Pre-amplifier internal input (BNC terminal) |
50 m (m/s 2)/DIV to 10 k (m/s 2)/DIV, 12 ranges x 6 types, Measurement resolution is 1/80 to 1/32 of range (change according to measurement sensitivity), Measurement sensitivity: 0.1 pC to 10 pC/ (m/s 2), Pre-amplifier internal input measurement sensitivity: 0.1 mV to 10 mV/ (m/s 2)
* When attached to recorder where full-scale (f.s.) = 20 divisions
100 m (m/s 2)/DIV to 20 k (m/s 2)/DIV, 12 ranges x 6 types, 1/160 to 1/64 of range, Measurement sensitivity: 0.1 pC to 10 pC/ (m/s 2), Pre-amplifier internal input measurement sensitivity: 0.1 mV to 10 mV/ (m/s 2)
* When attached to recorder where full-scale (f.s.) = 10 divisions
Amplitude accuracy: ¡¾2 % f.s. Frequency characteristics: 1 to 50 kHz +1/-3 dB
Low-pass filter: 500/5 kHz Pre-amplifier driving power supply: 2 mA ¡¾20%, +15 V ¡¾5%
Highest input charge : ¡¾500 pC (high sensitivity side 6 ranges), ¡¾50000 pC (low sensitivity side 6 ranges) |
Measurement ranges
Voltage input (BNC terminal) |
500 µV to 2 V/division, 12 ranges, Measurement resolution is 1/80 to 1/32 of range (change according to measurement sensitivity) * When attached to recorder where full-scale (f.s.) = 20 divisions 1 mV to 5 V/division, 12 ranges, 1/160 to 1/64 of range * When attached to recorder where full-scale (f.s.) = 10 divisions DC amplitude accuracy: ¡¾0.4 % f.s. Frequency characteristics: DC to 400 kHz +1/-3 dB Low-pass filter: 5/500/5 k/100 kHz Input coupling: DC, AC, GND Max. allowable input: 30 V rms or 60 V DC |
Common functions to all measurement ranges |
Anti-aliasing filter: can be turned ON or OFF, cut off frequency is the time axis, connected and set automatically by the frequency axis range |
Maximum sampling rate |
1 MS/s (simultaneous sampling of two channels) |
Max. rated voltage to earth |
30 V rms or 60 V DC (upper voltage which when applied to input channel casing or between input channels does not damage them) |
Accessories |
None * The input cord is optional |
|
|
|
|
|
|
|
|
|
|
|